Page 52 - ARNM-2-3
P. 52

Advances in Radiotherapy
            & Nuclear Medicine                                                  Image-guided interventional radiotherapy




            A                        B                       C                       D









            E                        F                       G                       H











            Figure 8. CT-simulator-guided radioactive seed implantation. (A) Patient setup and CT scan; (B) Pre-planning; (C) Repositioning 3D-PNCT or 3D-PCT;
            (D) Verification of seed needles position; (E) Confirming alignment between the seed needle position and pre-plan; (F) 125I seeds implantation; (G) Post-
            plan; (H) Post-operative care. Image provided by author.
            Abbreviations: 3D-PCT: 3D-printing coplanar templates; 3D-PNCT: 3D-printing non-coplanar template; CT-sim: Computed tomography simulator.

              The procedures of CT-sim guided radioactive seed   peer-review process conducted for this paper, directly or
            implantation are as follows: (i) patient positioning setup,   indirectly. Separately, other authors declared that they
            and contrast-enhanced CT-sim scan; (ii) transfer of CT-sim   have no known competing financial interests or personal
            images to the B-TPS for designing pre-plan; (iii) accurate   relationships that could have influenced the work reported
            repositioning of 3D-PNCT or 3D-PCT; (iv) insertion   in this paper.
            of seed needles; (v) CT-sim scan to confirm alignment
            between seed needle position and pre-plan; (vi) 125I seeds   Author contributions
            implantation; (vii) CT-sim scan to assess the displacement   Conceptualization: Junjie Wang, Ping Jiang
            of the 125I seeds, and supplementation of additional seeds,   Writing – original draft: Qiman Han, Yi Chen
            if necessary; and (viii) transferring patients back toward   Writing – review & editing: Junjie Wang, Zhe Ji, Yi Chen,
            (Figures 7 and 8).                                    Bin Qiu

            6. Conclusion                                      Ethics approval and consent to participate
            The advent of CT-sim has ushered in an era where   Not applicable.
            interventional   brachytherapy  has  attained  swift
            advancements, bringing forth the establishment of   Consent for publication
            standardized and harmonized technical systems that lay a   Patients gave consent to publish their images in this study.
            concrete foundation for a revolutionary transformation in
            tumor diagnosis and RT.                            Availability of data

            Acknowledgments                                    Not applicable.
            None.                                              References

            Funding                                            1.   Hwang EJ, Park CM, Yoon SH, Lim HJ, Goo JM. Risk factors
                                                                  for  haemoptysis  after  percutaneous  transthoracic  needle
            None.                                                 biopsies in 4,172 cases: Focusing on the effects of enlarged main
                                                                  pulmonary artery diameter. Eur Radiol. 2018;28:1410-1419.
            Conflict of interest                                  doi: 10.1007/s00330-017-5101-8
            Junjie Wang and Ping Jiang are the Editor-in-Chief and   2.   Kawashima H, Ohno E, Ishikawa T,  et al. Endoscopic
            the Editorial Board Member of the journal, respectively,   management of perihilar cholangiocarcinoma. Dig Endosc.
            but were not in any way involved in the editorial and   2022;34:1147-1156.


            Volume 2 Issue 3 (2024)                         7                              doi: 10.36922/arnm.3781