Page 45 - BH-1-2
P. 45

Brain & Heart                                                  RDN guided by novel blood pressure responses



            energy was 8–10 W for 60 s and increased to 12 W for   compare BP and heart rate at baseline, 5 – 7 days, and 6
            60 s if the SBP elevation was still evoked by the post-  – 2 months. The other measurement data were compared
            ablation electrical stimulation. The selection of exact   by Student’s t-test.  P  < 0.05 was considered statistically
            ablative powers depended on the impedance. The ablation   significant. Data analysis was performed by SPSS 23.0
            endpoint was defined as the disappearance or significant   software (IBM Corporation, Armonk, New York).
            blunting of the SBP-elevated response during the post-
            ablation electrical stimulation. After the completion of all   3. Results
            intra-arterial high-frequency electrical stimulation and   A total of 14 patients who received selective RDN guided
            radiofrequency ablation, bilateral renal arteriography was   by RNS were included in this study. The patients were
            performed repeatedly to evaluate for abnormalities such as   relatively young (43 ± 9 years old), mostly male (11/14),
            dissection, perforation, thrombosis, and other procedure-  and not lean (BMI: 26.1 ± 2.2 kg/m ), with an ABP of 141 ±
                                                                                          2
            related adverse events.                            12/94 ± 8 mmHg and an OBP of 158 ± 24/101 ± 19 mmHg
            2.3. Measurement                                   at baseline. The remaining demographic data and clinical
                                                               characteristics are shown in Table 1.
            The amplitude of beat-to-beat BP during the procedure
            can be measured and stored automatically by the Lead   The beat-to-beat SBPs at 260 RNS sites before RDN
            Multi-channel Physiological Analysis software (Sichuan   were obtained to analyze. The following 5 types of BP
            Jinjiang Electronic Science and Technology Limited,   responses were observed (Figure 2 and Table 2) and are
            Chengdu, China). It could be regarded as unchanged when   further elaborated below:
            SBP fluctuation is in the vicinity (in the range of less than   i.   Type 1: BP persistently elevated above baseline.
            5 mmHg) of baseline, based on the lower beat-to-beat SBP      SBP in 74 sites (28.5%) continued to elevate during
            variability (1.4–2.8 mmHg) with an invasive monitoring   RNS. The peak SBP during the stimulation increased
            technique . We used the intraoperative BP data observed   significantly and exceeded the baseline by 14.3 ±
                    [20]
            in an animal study  to verify whether these possible trend   8.9 mmHg. The terminal SBP of the stimulation was
                          [18]
            curves exist in humans. A quantitative method for judging   still higher than the baseline by 11.6 ± 7.9 mmHg.
            the trend of BP change during electrical stimulation was
            used  in this study.  According to  the change of  specific   ii.  Type  2: BP dropped and then elevated above the
            values, such as the maximum, minimum, and end values   baseline.
            during the stimulation from the pre-stimulation value,      SBP in 47 sites (18.1%) dropped initially, then recovered
            the fluctuating trend of an SBP curve can be preliminarily   and exceeded the baseline during RNS. The valley SBP
            determined. Thus, in this study, using 10 cardiac cycles   dropped markedly, was lower than the baseline by
            as a unit of measurement, we first identified the specific   14.0 ± 7.1 mmHg, and then recovered. The peak SBP
            values containing the minimum SBP, the maximum SBP,   increased significantly and exceeded the baseline by
            the end-stimulation SBP, and their occurrence times. Then   13.2 ± 7.6 mmHg. The terminal SBP of the stimulation
            we calculated the SBP change of these specific values from   was still above the baseline by 8.3 ± 9.6 mmHg.
            the pre-stimulation SBP to show and classify the trend of
            BP at each stimulation site (a detailed process is shown   iii.  Type  3:  BP  dropped  and  then  recovered  but  never
            in Figure 1). ABP, average heart rate from the report of ABP   went over the baseline.
            monitoring, and anti-hypertensive medication regimens at      SBP  in 50  sites  (19.2%)  dropped initially  and then
            baseline, 5 – 7 days, and 6 – 12 months were collected.  recovered but never went over the baseline during
            2.4. Statistical analysis                             RNS. The valley SBP dropped markedly and was lower
                                                                  than the baseline by 16.7 ± 8.4 mmHg, and the peak
            Categorical variables were expressed as frequency or   SBP recovered but never exceeded the baseline (−1.8 ±
            percentage and statistically compared using the chi-square   5.9 mmHg). Terminal SBP was higher than the valley
            or Fisher’s exact tests. The Shapiro-Wilk test was used to   SBP by 13.1 ± 7.0 mmHg, which was -3.6 ± 5.6 mmHg
            determine whether the measurement data had a normal   of the baseline.
            distribution. P > 0.05 was considered to conform to the
            normal distribution, and then the measurement data were   iv.  Type 4: BP fluctuated in the vicinity of baseline.
            expressed as the mean ± standard deviation. Non-normally      SBP in 71 sites (27.3%) showed no observable change
            distributed data were expressed as a median with the   during RNS. BP fluctuated in the vicinity of the
            interquartile range (IQR) (25% IQR, 75% IQR). Univariate   baseline (−1.0 ± 2.1  mmHg at the valley and 1.3 ±
            repeated  measurement  analysis  of  variance  was  used  to   2.0 mmHg at the peak) during electrical stimulation.


            Volume 1 Issue 2 (2023)                         3                         https://doi.org/10.36922/bh.0384
   40   41   42   43   44   45   46   47   48   49   50