Page 102 - AN-3-4
P. 102

Advanced Neurology                                              Shuntogram technique in programmable valves



            patency. Ideally, the initial images should display only   then every 5 min, as detailed above. The X-ray tube and
            this small tracer volume within the reservoir of the shunt,   plate can be adjusted caudally as needed to keep the distal
            with no proximal or distal flow, although proximal flow   injectate within view.
            may occur in low-pressure systems. The initial image
            review also confirms that the injectate is within the   3. Results and discussion
            shunt lumen and not in subcutaneous tissue. If no distal   Table 1 summarizes the characteristics of the 25 publications
            flow is observed within the first 5 – 10 min, the patient   discussing shuntogram techniques. These studies exhibit
            is then placed upright in a chair or allowed to stand for   considerable  variability  in  areas  such  as  patient  age,
            5 min before re-imaging in the supine position to assess   clinical indication, positioning, skin preparation, duration,
            for siphoning-related flow. If the flow is still absent and   contrast agents, imaging protocols, patency assessment,
            a programmable valve is present, the valve should be   diagnostic criteria (negative and positive), adjustments
            re-read to confirm its setting and reprogrammed to the   to shunt flow following positive failure, and measures of
            lowest setting if not already set there. Absence of flow after   sensitivity and specificity or result/clearance times. Further
            these steps indicates an obstruction. As a final adjunctive   details are available in Table S1.
            step, the shunt reservoir can be manually pumped to   In  the  subsequent  discussion,  we  chronologically
            determine whether the obstruction is partial (flow occurs   examine  the  evolution  of  shuntogram  techniques  from
            under elevated pressure) or complete (no flow despite   1981 to 2024.
            pumping). This final step disturbs the steady state of the                     9
            shunt system and is not used to rule out shunt failure,   In 1981, French and Swanson published a study on
            but it helps determine if the system remains in continuity   radionuclide imaging shuntography to evaluate shunt
            (Figure 2).                                        patency. They  conducted  78 shuntograms in  patients
                                                               exhibiting clinical features of obstruction. Patients were
              When iodinated contrast is used rather than nuclear   placed in a supine position, and scalp preparation involved
            medicine agents, images are obtained using static X-rays   shaving and disinfecting with betadine. A 23-gauge needle
            of the skull and abdomen. We recommend starting with   was  inserted  through  a  small  incision  in  the  scalp  into
            a lateral or oblique view in the sagittal plane aligned with   the reservoir, and   99m Tc-pertechnetate or  In-DTPA was
                                                                                                111
            the shunt valve (Figure 3). X-rays are taken at 0 min and   injected. Over 100,000 images  were obtained using the



































                              Figure 2. Flowsheet for troubleshooting and diagnosing impaired contrast flow on shuntogram
                                                  Abbreviation: CSF: Cerebrospinal.


            Volume 3 Issue 4 (2024)                         4                                doi: 10.36922/an.4180
   97   98   99   100   101   102   103   104   105   106   107