Page 24 - manuscript_ijb05583
P. 24

While deposition with the chip proved precise, switching to a dedicated line for cell delivery

               put  much  less  stress  on  the  microfluidic  device.  Given  its  lengthy  fabrication  process,
               preserving  its  functionality  was  a  clear  benefit.  Furthermore,  the  dedicated  line  allowed  a

               higher fluid pressure and thereby a higher dispensing throughput (Error! Reference source
               not found.). Given that liquid dispensing was required twice per culture well – once for pre-

               filling and once for cell delivery –, the speed of liquid delivery was a major factor in liquid

               dispensing.
               Moreover, using the microfluidic device itself for dispensing would have resulted in a wasted

                                                           6
               volume of 20 mL of medium containing 4x10  cells for a 384-well plate, as cell-laden medium
               would constantly flow to the waste channel to keep a constant flow rate while the dispensing

               step is in progress. For the 1536-well plates, however, cell delivery via a dedicated line proved

               challenging. First, the well dimensions were 1.7 x 1.7 mm², requiring the delivery line with an
               outer diameter of 1.6 mm  to  be perfectly centered. While this  was alleviated by adding  a

               dispensing tip to the end of the tubing, dislodging the droplet from the needle proved more
               problematic, as the working volume of a 1536-well plate was specified with 4 – 12 µL. Initially,

               the needle was kept so close to the well-plate that moving to the next well wiped the tip.
               However, as the tip was not always positioned at the same height, it would get entangled when

               moving the well plate, resulting in double dispensing and overflowing wells. In addition to

               overflowing wells, wiping the nozzle between wells increased the risk of infection. For the
               1536-well plates, cell deposition using the microfluidic device was therefore favored, as having

               an  air-pressurized  outlet  allowed  to  precisely  eject  even  small  volumes  down  to  10  µL
               repeatedly.


               Excluding drying and UV sterilization, the automated workflow for preparing a fully seeded

               plate required approximately 20–25 min for 96 wells, 1 h 15 min for 384 wells, and about 5 h
               for 1536  wells. However, manual steps still added substantial  time to  the overall process,

               particularly  the  drying  step,  which  took  several  hours.  This  step  could  be  significantly
               shortened in the future by replacing heat evaporation with direct liquid removal or vacuum-

               assisted drying. Additionally, parallelization of plate handling or the use of alternative carrier
               liquids such as alcohols, which evaporate faster and maintain sterile conditions, could have

               further reduced preparation time. Of note, microscaffolds could have been presorted and stored

               in the liquid, allowing the cell culture workflow to start when needed. Importantly, the design
               of the device enabled a clear path toward scalability, as it allows to multiplex multiple units in






                                                           23
   19   20   21   22   23   24   25   26   27   28   29