Page 275 - IJB-10-1
P. 275

International Journal of Bioprinting                                    Permeability of NiTi gyroid scaffolds











































                                        Figure 7. Contour plot with color-coded permeability coefficient.
            LPBF demonstrate a continuous and smooth near-circular   morphology can decrease the velocity of the fluid. Moreover,
            front of fluid in the  x-y  plane, providing high-accuracy   surface roughness affects wetting properties in unsaturated
            measurements. Finally, the method allows the exclusion of   measurements, leading to lower permeability values in
            finite-size effects due to the unlimited number of in-plane   comparison to a hydrophilic surface. Third, according
            cells for the prescribed plane.                    to the results of the micro-CT analysis, all scaffolds
                                                               demonstrated positive deviation from the prescribed
               Calculated and experimentally obtained permeability
            coefficient values are presented in  Figure 8 as a function   geometry; consequently, the apparent density of the samples
                                                               was  higher,  with  the  identified  dependence  of  relative
            of porosity for comparison with literature results. The   geometry deviation from the unit cell size via micro-CT
            experimentally obtained point corresponds to the mean value   analysis demonstrated in Figure 3d. All these factors were
            of 3 measurements with indicated standard deviation. The   considered to be responsible for the slope difference between
            same near-linear trend in permeability change is observed   the simulation and experimental results.
            for the experimental and CFD model, with numerics
            returning values about three times higher. The absolute   Permeability coefficients for real bones and different
            difference in values is low for the higher apparent densities   porous  scaffolds  involving  other  types  of  unit  cells  are
            and grows with an increase in porosity. A difference of 31%   presented in  Table 4. It shows that real bones exhibit
            was observed for sample 254, and 77% for sample 203.   a wide range of permeability values. The biomimetic

               The difference between the simulation and experiment   design of a scaffold requires a full adaptation not only
            can be attributed to several factors. First, only laminar   of the permeability range but also of the pore size and
            flow was simulated in the FE model. It was experimentally   porosity. On that note, the controllability of the design
            demonstrated  in  the  study  by  Bobbert  et al.   that  the   is an advantage of the additive manufacturing approach
                                                  35
            permeability is significantly higher in the case of laminar   for the implementation of patient-specific implants. In
            flow  in  comparison  with  turbulent  flow.  Second,  surface   existing works, relatively low  and high 17,36  permeability
                                                                                      9,35
            condition plays an important role in the resulting wall   coefficient values are presented. This study demonstrated
            shear stress (WSS). Thus, the demonstrated channel  wall   via FE simulation and experimental measurements that


            Volume 10 Issue 1 (2024)                       267                         https://doi.org/10.36922/ijb.0119
   270   271   272   273   274   275   276   277   278   279   280