Page 76 - IJB-8-1
P. 76

AJ P of Bioelectrical Devices
                                                                                             ®
           A                                                   A










            B










           Figure 8. Printing of the device: on (A) glass substrate and (B)
           NTE substrate.                                      B

           for a longer period of time (2 h against 8 min) at 150°C.
           The results confirmed the previous observations, showing
           a drastic cytotoxic ink behavior already after 24 h, with
           no value  detected  at the following points of 48  h and
           120 h (Figure 7D). As such, the printed PEDOT: PSS ink
           is considered cytotoxic for NSCs.
               To verify whether this ink behavior is related to a
           dose-dependent  and/or sintering condition,  additional
           indirect  cell  viability  assays were carried  out on
           nanometric spin-coated PEDOT: PSS samples, and at six
           different curing conditions (150°C for 8, 60, and 120 min,   Figure  9.  Characterization  of  the  device:  (A)  Contour
           and 200°C for 8, 60, and 120 min). Figure 7E indicates   profile  of  printed  electrodes  at  different  printing  layers
           that non-significant variations were detected among the   n = (5, 20, 30, 40, 50), showing the potentiality of reaching ca. 30 µm with n 40;
           different annealing conditions after 48 h. The conditions at   (B)  Electrical  impedance  of the printed  circuit  (40 layers) on
                                                               Parylene-C NTE substrate. Impedance (Ω) and phase (°) are plotted
           150°C and 120 min were then taken as reference settings   versus frequency (Hz).
           for the preparation of the next spin-coated samples, and
           the ATP was tested up to 120 h. The results, shown in   electrodes at different printed layers, demonstrating the
           Figure 7F, confirm the hypothesis of a dose-dependent   possibility to reach the same height of the micro-channels,
           behaviour of the ink. Despite a significant reduction of the   ca. 30 µm, with the number of layers n ≥ 40. Electrodes
           cellular viability at 48 h with respect to the plastic control
           (40.61%  less),  the  biocompatibility  of  the  nanometric   of the same height are indeed important for establishing
           spin-coated films was considered acceptable, resulting in   a  constant  electrical  field  that  is  well  aligned  with  the
           29.38% less biocompatibility than the plastic control at   channel  direction.  In  addition,  Figure  9B  depicts  the
           120 h.                                              electrical impedance of the printed device, showing that
                                                               the behavior of the circuit correctly follows a theoretical
           3.3. Target application                             RC circuit. In particular, the impedance was detected in
                                                               the range of 1 – 2 kΩ at 1 kHz, in accordance with typical
           The  printing  strategy  depicted  in  Figure  5C  was   values of neural electrodes for detecting neural activity,
           ultimately  applied  to  realize  the  target  application.   demonstrating the validity of the device.
           Figure  8A and B show  images of the prototypes
           fabricated with AJ P printed interconnects and electrodes   4. Conclusions
                          ®
           on  glass and  Parylene-C-coated  NTE  substrates,
           respectively.  Each  electrode  was  ca.  33  µm  thick  and   In NTE, multiple  and complementary  guidance  cues,
           ca. 2 mm wide (including overspray), with an electrical   such as topographical,  electrical,  mechanical  and/or
           resistance, R = 16 Ω. Figure 9 reports a characterization   biochemical  ones, are typically  applied on a neural
           of the device  printed  on NTE substrates. In particular,   scaffold  to  regulate  neural  cell  activities  for  in  vitro
           Figure  9A  displays  the  contour  profile  of AJ  printed   studies.  This work concerns the potential  of  Aerosol
                                                   ®
           62                          International Journal of Bioprinting (2022)–Volume 8, Issue 1
   71   72   73   74   75   76   77   78   79   80   81