Page 165 - IJB-8-3
P. 165

Hunag, et al.
           daily motions of piano-playing, pen-writing, and can-  medium, and smallest phalanges were ranged from
           opening.  The test setup was needed to ensure joint   30% to 72%, of which 72% occurs at the metaphyseal
           surface alignment and full contact (Figure  4C). The   of the largest phalanx size (Table 2). In addition, there
           static dislocation test was performed by applying   were 5 PIP joint implant combination sets, that is,
           vertical downward compressive force at a constant   0% (articular surface) to 0% (stem), 10% (articular
           speed of 5 mm/min on the implant, which was stopped   surface) to 0% (stem), 10% (articular surface) to 10%
           when the load dropped or the joint surface became   (stem), 20% (articular surface) to 10% (stem), and 20%
           misaligned.  The maximal dislocation force and the   (articular surface) to 20%  (stem) that  can be placed
           damage situation were recorded and compared with the   into  this  largest  phalanx.  Finally,  the  worst  structure
           results from the previous literatures.              case, that is, weakest structure in all set combinations
               Regarding the dynamic test, cyclic force sine wave   was the enlarged 20% of the articular surface and 10%
           mode at 1  Hz frequency was performed for a total of   of the stem size due to the articular surface level was
           25,000 cycles to simulate the normal joint activity 1 month   larger than the corresponding stem level combination to
           after the operation . The force was initiated at 50% of   produce a larger force state (Table 2).
                          [13]
           the maximum dislocation force in the corresponding three   Figure 5A shows the PIP joint implant including
           static tests. Force was then increased at a gradient of 10%   3D printing proximal phalanx with polished articular
           until the cyclic load limit number was obtained maximum   surface and middle phalanx consisting of 3D printing
           25,000 cycles (cannot pass 25,000 cycles). Three samples   cone stem and milling UHMWPE curved articular
           from each set were used to obtain the corresponding   surface connected by hook mechanism.  The major
           percentages for three static loads with different angles.   axis and minor axis of the metaphyseal and diaphyseal
           The misalignments were then observed and recorded.  ellipses and the total length of the stem for proximal and
                                                               middle phalanxes were recorded. The error percentages
           3. Results                                          were all within ±5% compared with the CAD design

           The medullary cavity ratio to the cross-sectional area   value (Tables 3 and 4).
           at the phalanx stem top and bottom sides of largest,    The maximum  pull-out force of the proximal
                                                               phalanx prosthesis was 727.8N ± 45.6N (average value
           A                            B                      ± standard deviation) and the loosening failure mode is
                                                               shown in Figure 4B. The maximum connection force of
                                                               the hook mechanism to cone stem of the middle phalanx
                                                               is 49.9N ± 2.0N (average value ± standard deviation) and
                                                               the  hook mechanism  was broken instead  of cone  stem
                                                               pull-out from middle phalanx in Figure 3D.
                                                                   The dislocation force of 25°, 35°, and 55° for
                                                               PIP joint implant were 525.3N ± 21.2N, 316.0N±
                                                               17.2N, and 115.0N ± 1.8N, respectively (Table 5) and
           C                                                   Figure  6 shows the load displacement  diagram of all
                                                               dislocation tests.  Figure  4D shows the dislocation
                                                               situation  under  35°  load  condition  and fractured  bone
                                                               around the UHMWPE articular surface of the middle
                                                               phalanx was found. Corresponding percentages of three
                                                               static  loads  with  different  angles  that  can  pass  25000
                                                               dynamic cyclic loads were 50% (26.25~262.5N), 50%
                                                               (15.8~158N), and 80% (9.2~92N) for 25°, 35°, and
                                                               55° load conditions, respectively (Table 6). Figure 6E
                D                E                             shows the misalignment of the PIP joint implant under
                                                               55° dynamic load cycles.
                                                               4. Discussion

                                                               Although PIP joint  arthroplasty is well-established,  it
                                                               can still be a challenging task for surgeons [1,2] . Current
                                                               commercial artificial PIP joint implants have the following
                                                               limitations; one-piece silicone implants lack appropriate
           Figure 4. Tests of anti-dislocation ability for PIP joint prosthesis at   rigidity, resulting  in lateral  instability  and weak pinch,
           25°, 35°, 55°and on failure mode.                   and implant breakage due to repetitive usage [1-4] . Although

                                       International Journal of Bioprinting (2022)–Volume 8, Issue 3       157
   160   161   162   163   164   165   166   167   168   169   170