Page 221 - IJB-9-4
P. 221

International Journal of Bioprinting                Simulation-based comparative analysis of nozzles for bioprinting



































                           Figure 3. Outlet pressure (Pa) of pneumatic and piston-driven simulations and the two analyzed nozzles.

            Our validation tests recreated the experimental validation   printing nozzle. It is not possible to generalize the results to
            performed  by  Liravi  et al. .  The  extruded  strand  was   other conical tip gauges and nozzle output diameters with
                                  [47]
            measured and compared with the simulation results, using   the current data. The internal geometry of the conical tips
            the total height and the maximum width of this extruded   remains constant for gauges smaller than 27G, but some
            strand. This indirect validation of our simulations was used   of the internal lengths of the E3D V6.4 3D printing nozzle
            due to the small outlet diameter of the Cone and Nozzle   vary with each output diameter .
                                                                                        [53]
            that was nearly impossible to measure the pressure or the
            velocity without disturbing the normal flow and thus modify   3.1. Outlet and inlet pressure
            the results. A Sony Alpha 7 ii (ILCE-7m2) camera with a   Figure 3 shows the outlet pressure measured at end of
            Sony f 2.8/90 OSS macro was used to record the extruded   the Cone and Nozzle  for pneumatic and piston-driven
            Cellink Bioink strand with a 1920 × 1080 pixel resolution   simulations.
            and 0.5× magnification. The distance between the camera   Analyzing the pressure of pneumatic simulations,
            lens and the nozzle tip was 28  cm (focal distance). The   using 15 kPa as the simulation inlet, the pressure evolution
            microextrusion was performed in Cellink BioX bioprinter   in both geometries is similar, showing a low-pressure peak
            with a preset of 15 kPa for inlet pressure. Three videos for   which is explained by the formation and falling of a drop,
            each of the four experimental tests were recorded, from   as detected in previous works . Therefore, the maximum
                                                                                      [50]
            which frames were automatically extracted using Open CV   outlet pressure before the low-pressure peak reaches 2.75
            and analyzed using ImageJ 1.53e software.          and 1.54 kPa for the Nozzle and Cone, respectively. The
                                                               outlet pressure of the Nozzle is approximately twice that
            3. Results and discussion                          of the Cone. Regarding timing, the peaks are in 0.586 and
                                                               6.860 s for Nozzle and Cone, respectively. The difference
            In this section, results obtained for the three analyzed   in times might be caused by the total amount of extruded
            parameters are presented. However, since later     bioink. In this sense,  Figure 4 shows that the flow rate
            accumulation of material is not produced in a real   of the Nozzle geometry is higher than that of the Cone.
            bioprinting procedure because of the X–Y extruder head   Specifically, the drop volume extruded by the Nozzle
            movement, data are only analyzed until the drop falls,   is  12.70  mm ,  while  the  one  by  the  Cone  is  9.72  mm
                                                                         3
                                                                                                             3
            although figures show the full 10 s simulations.
                                                               through the same cross-sectional area. So, the Nozzle
               Additionally, the results presented in this section only   geometry extrudes  approximately  15  times  more bioink
            refer to the 22G conical tip and the 0.4 mm E3D V6.4 3D   than the Cone (21.67 vs 1.42 mm /s). This large difference
                                                                                          3
            Volume 9 Issue 4 (2023)                        213                         https://doi.org/10.18063/ijb.730
   216   217   218   219   220   221   222   223   224   225   226