Advances in Radiotherapy
            & Nuclear Medicine                                                     Optimal tube voltage for dental CBCT



            also the mean energy and amount of released photons in   Availability of data
            an X-ray beam. 3
                                                               Data are available from the corresponding author upon
            4. Conclusion                                      reasonable request.

            In the present study, we identified the optimal exposure   References
            protocols for the Planmeca ProMax 3D Mid CBCT unit   1.   Venkatesh E, Elluru SV. Cone beam computed tomography:
            (Planmeca, Helsinki, Finland). The highest selectable tube   Basics and applications in dentistry.  J  Istanbul  Univ  Fac
            current at constant tube voltage was revealed to be the   Dent. 2017;5:102-121.
            most efficient in noise reduction. However, an increase
            in noise of about 0.09% was noticed at the same exposure      doi: 10.17096/jiufd.00289
            setting when the tube voltage was varied at constant mAs,   2.   Lemoigne Y, Caner A, Rahal G.  Physics for Medical
            given that noise increases at lower mAs. Hence, to achieve   Imaging Applications. vol.  240. Netherlands: Springer
            lower noise with the higher dose, 70 kV and 168.16 mAs   Dordrecht; 2007.
            should be applied.                                 3.   Yel I, Booz C, Albrecht MH, et al. Optimization of image
              In conclusion, low noise protocols should consist of a   quality and radiation dose using different cone-beam CT
            reduction in the tube current rather than a reduction in   exposure parameters. Eur J Radiol. 2019;116:68-75.
            kVp, as the former would result in a smaller level of noise      doi: 10.1016/j.ejrad.2019.04.005
            and consequently less quality degradation in the image.  4.   Panmekiate S, Rungwittayathon P, Suptaweeponboon W,
              The X-ray tube’s potential, tube current, and exposure   Tangtraitham  N,  Pauwels R.  Optimization  of  exposure
            time may influence the radiation production, image    parameters in dental cone beam computed tomography
                                                                  using a 3-step approach.  Oral Surg Oral Med Oral Pathol
            quality, and dose produced. Hence, tube  performance   Oral Radiol. 2018;126(6):545-552.
            should be assessed in a series of periodic tests.
                                                                  doi: 10.1016/j.oooo.2018.08.004
            Acknowledgments                                    5.   Bamba J, Araki K, Endo A, Okano T. Image quality
            The authors would like to thank all staff at the Imaging Unit   assessment of three cone beam CT machines using the
            of AMDI, Universiti Sains Malaysia, Penang, Malaysia for   SEDENTEXCT  CT  phantom.  Dentomaxillofac Radiol.
            their help and support throughout this work.          2013;42(8):20120445.
                                                                  doi: 10.1259/dmfr.20120445
            Funding                                            6.   Pauwels R, Seynaeve L, Henriques JC, et al. Optimization
            None.                                                 of dental CBCT exposures through mAs reduction.
                                                                  Dentomaxillofac Radiol. 2015;44(9):20150108.
            Conflict of interest                                  doi: 10.1259/dmfr.20150108

            The authors declare that they have no competing interests.  7.   De Oliveira MVL, Santos AC, Paulo G, Campos PSF,
                                                                  Santos J. Application of a newly developed software program
            Author contributions                                  for image quality assessment in conebeam computed
            Conceptualization: James  Anthony  Rabba,  Noor  Diyana   tomography. Imaging Sci Dent. 2017;47(2):75-86.
               Osman                                              doi: 10.5624/isd.2017.47.2.75
            Investigation: James Anthony Rabba, Hanis Arina Jaafra   8.   Park HN, Min CK, Kim KA, Koh KJ. Optimization of
            Methodology: James Anthony Rabba, Noor Diyana Osman,   exposure parameters and relationship between subjective
               Godwin Irinam Efenji                               and technical image quality in cone-beam computed
            Writing – original draft: James Anthony Rabba         tomography. Imaging Sci Dent. 2019;49(2):139-151.
            Writing – review & editing: Noor Diyana Osman, Fatanah      doi: 10.5624/isd.2019.49.2.139
               Mohamad Suhaimi, Felix Omachoko Uloko, Dauda
               Biodun Amuda, James Anthony Rabba               9.   Alawaji Y, MacDonald DS, Giannelis G, Ford NL.
            Ethics approval and consent to participate            Optimization of cone beam computed tomography
                                                                  image quality in implant dentistry.  Clin Exp Dent Res.
            Not applicable.                                       2018;4(6):268-278.
                                                                  doi: 10.1002/cre2.141
            Consent for publication
                                                               10.  Brasil DM, Pauwels R, Coucke W, Haiter-Neto F,
            Not applicable.                                       Jacobs  R. Image quality optimization of narrow detector


            Volume 2 Issue 2 (2024)                         6                              doi: 10.36922/arnm.2972
   61   62   63   64   65   66   67   68   69   70   71