Page 80 - BH-2-1
P. 80

Brain & Heart                                                               Pacing strategy for heart block



                                                                A                          C















                                                                B
            Figure 1. Baseline electrocardiogram indicating complete right bundle
            branch block.

            into the right ventricle. A SelectSecure lead (Model 3830;
            Medtronic, Inc., Minnesota, USA) was then screwed
            into the RV septum. The pacing lead was maneuvered to
            the tagged area, in which unipolar electrode tip pacing
            demonstrated QRS complex with a “W pattern” in ECG
            lead V  and RS pattern in the inferior lead. As the lead was
                 1
            advanced, the W shape in lead V  changed to a QR pattern
                                      1
            with a QRS duration of 158 ms (Figure  2). Notably, no
            discernible LBB potential was recorded. Threshold testing,   Figure  2.  (A and B) X-ray images and (C) paced QRS during left
            utilizing unipolar pacing, revealed left ventricular septum   ventricular septal pacing.
            pacing (LVSP) without LBB capture, consistent with the   Abbreviations: LAO: Left anterior oblique; LVS: Left ventricular septum;
            criteria established by Wu  et al.  The pacing parameters   RAO: Right anterior oblique.
                                      2
            were within acceptable ranges (pacemaker sensing:
            10 mV; pacing threshold: 0.4 mV/0.4 ms, and pacing   2.0 V/0.4 ms, which resulted in favorable cardiac electrical
            impedance: 580  Ω). An unexpectedly narrower QRS of   and mechanical synchrony (Table 1 and Figure 3).
            non-RBBB morphology was noticed when a bipolar pacing   In comparison with LVSP, BSP led to a pronounced
            configuration was  adopted. This  finding  indicated that   reduction in QRS duration. However, no R’ was observed
            BSP was achieved with the capture of both aspects of the   in ECG lead V  (QRS with RBBB pattern was not observed)
                                                                          1
            septum when bipolar pacing was performed.          during BSP (Figure  4). BSP improved the IVMD when
              ECG measurements were obtained using a multichannel   compared with LVSP (Figure 5).
            recorder. Echocardiography was performed 1 week after the
            pacemaker implantation procedure to evaluate mechanical   3. Discussion
            synchrony. The interventricular mechanical delay (IVMD)   The present case report is the first case description of BSP,
            was measured as the time interval between the onset   showcasing a reduction in IVMD time compared to the
            of QRS and the beginning of systolic waves of aortic   LVSP. This unexpected achievement was realized through
            and pulmonary ejections, using conventional Doppler.    the right septal capture with an LVS lead. Our findings
                                                          3
            Intraventricular dyssynchrony was evaluated using a   underscore the potential of BSP to attain optimal cardiac
            Yu index, defined as the standard deviation of the time   electrical and mechanical synchrony.
            between the onset of QRS and the peak systolic velocity   A notable complication frequently associated with
            of tissue Doppler for 12 left ventricular (LV) segments   TAVR is heart block, especially prevalent in patients with
            (six basal and six middle) in apical triplane-mode (using a   prior RBBB.  As documented in several studies, chronic
                                                                         4
            4-Vprobe).  Septal-posterior wall motion delay (SPWMD)   RV pacing is linked to an increased risk of heart failure.
                    3
                                                                                                             4
            was measured through M-mode as the distance between   Recently, LBBP has emerged as a new pacing approach.
                                                                                                             5
            the first maximum systolic inward motion of the septum   However, while LBBP produces a relatively narrow QRS
            and the maximum inward motion of the posterior wall.  complex with fast left ventricular activation, concerns
              Anodal capture using the ring electrode in the   persist regarding the presence of RBBB or even incomplete
            RV septum was observed using bipolar pacing at     RBBB during pacing, leading to potential interventricular


            Volume 2 Issue 1 (2024)                         2                         https://doi.org/10.36922/bh.1670
   75   76   77   78   79   80   81   82   83   84   85