Page 81 - IJB-2-1
P. 81

Ryan D.  Boehm, Panupong Jaipan,  Kai-Hung Yang,  et al

             0.52 ≥ S M −  2  ≥− 0.01, with a difference of means of   system procedure and  results from the  manufac-
                     1 M
            0.25. For the nine samples tested using each method in   turer-described  procedure for Day 0  –  Day 3. These
            this comparison, no statistical difference was observed   results from both procedures were well into the nega-
            at a 95%  confidence level between  the  microneedle   tive ratio values; although a large discrepancy in the
            sampling   system  procedure  results  and  the    results from the two procedures was noted, both pro-
            manufacturer- described procedure results. Although a   cedures  showed  that  the  samples  were  negative  for
            difference between the  microneedle sampling system   histamine at the 50 ppm threshold. The Day 5 results
            procedure results and the  manufacturer-described   showed a dip in the ratio readings. Both procedures
            procedure results was noted for the PBS-incubated   detected increasing levels of histamine; however, both
            samples, the histamine-spiked samples showed com-  procedures showed that the samples were negative for
            parable results for the  microneedle sampling system   histamine at the 50 ppm threshold. A substantial dip in
            procedure and the manufacturer-described procedure.   the  microneedle sampling system  procedure results
                                                     ®
            Figure 3  shows the mean values of Accuscan   ratio   was noted at Day 5. At the Day 7 time point, the de-
            readings for  each type of sample.  A decrease in the   crease in readings for both test procedures continued.
            relative differences between  the ratios of  the micro-  Using the confidence interval comparison method des-
            needle sampling system procedure results and the   cribed previously, no difference between the results
            manufacturer-described  procedure results was noted   from the two test  methods was noted at Day 7. The
            as the sample type changed from PBS-spiked to his-  results  remained  negative  at the 50  ppm histamine
            tamine-spiked.                                     detection level at all time points over the 7-day time
                                                               course.

















                                                          ®
            Figure 3.  Graph of the mean  (±  S.E.M.) of the  Accuscan    Figure 4.  Graph of the mean  (±  S.E.M.) of the  Accuscan
                                                                                                            ®
            reader test:control ratio data for histamine-spiked tuna samples   reader test:control ratio data for tuna samples over the refrige-
            acquired through the microneedle  sampling  system procedure   rated 7-day time course. Color intensity ratios acquired through
            and the  manufacturer-described  procedure. For comparisons   the microneedle sampling  system procedure and  the  manu-
            marked  with  “*”, no  statistical difference was  noted  between   facturer-described  procedure  were compared. For comparison
            the results from the microneedle sampling  system procedure   marked  with  “*”, no  statistical difference was  noted  between
            and the manufacturer-described procedure when looking at the   the microneedle sampling  system procedure and the  manufa-
            difference of means 95% confidence intervals for n = 9 of each   cturer-described  procedure when  looking  at the difference of
            test type.                                         means 95% confidence intervals for n = 5 of each test type.


            3.3 Testing of Tuna from Time Course and Spoiled     While the testing of time course samples remained
            Samples
                                                               negative, a spoiled piece of tuna was identified during
            In addition to the histamine-spiked tuna samples that   the study. As described earlier, pieces of tuna that
            were used to compare the microneedle sampling sys-  were incubated in PBS overnight under refrigerated
            tem procedure and the manufacturer-described proce-  conditions  were  meant to serve as negative controls
            dure, tuna samples that were left in a refrigerator for   during the histamine-spiking tests. However, one of
            seven days were tested using both procedures. Figure   the pieces of tuna that was tested in this  manner re-
            4 shows the test:control ratio values and standard error   turned  a  positive result for  histamine. It  was deter-
            of the  mean values  for each set of  measurements at   mined that the tuna sample was spoiled upon acquisi-
            Day 0  –  Day  7 time  points. Large differences  were   tion  from  the  local  fresh  market.  This  spoiled  tuna
            noted between results from the microneedle sampling   sample provided an opportunity to compare the  mi-

                                        International Journal of Bioprinting (2016)–Volume 2, Issue 1      77
   76   77   78   79   80   81   82   83   84   85   86