Page 79 - IJB-2-1
P. 79

Ryan D.  Boehm, Panupong Jaipan,  Kai-Hung Yang,  et al

            was used  to  optically  measure  the intensities of the   creating (a) the microneedle arrays for sample acquisi-
            colored test and control lines in the lateral flow strip.   tion from the tuna samples, (b) the microneedle holder,
            Each test strip has colored control and test lines that   and (c) the lateral flow test holder. As seen in Figure
            appear upon introduction of a fluid sample. The ratio   2A–C, the microneedles were arranged in staggered 3
            of the color intensities of these lines was used to as-  × 3 rows in each  microneedle sampling system. The
            sess if a sample was positive or negative for histamine   trapezoidal eyelet extending from the microneedle
                                          ®
            at the 50 ppm level. The Accuscan  reader provided a   base toward  the  microneedle tip  served  as a capture
            ratiometric readout  of the lines, which  was  used to   point for fluid from the tuna sample. The  micro-
            differentiate between a positive result and a negative   needles had heights  of 1150 µm and base widths of
            result. If the ratio of the test:control lines was greater   950 µm.  Figure 2B–C  show the insertion of  the mi-
            than or equal to one, the test was read as negative for   croneedle array into a piece of tuna and the indention
            histamine. If the ratio of the test:control lines was less   left in  the tuna sample upon removal of the micro-
            than one, the test was read as positive for histamine.   needle array.
            For each  test lateral flow strip, three  measurements   As illustrated in Figure 1B–D, following insertion
            were taken in  short succession  with  the reader; the   of the microneedle array into the tuna sample, the mi-
            average ratio was used to determine positive versus   croneedle array was inserted into the central chamber
            negative for the individual strip. For the calibration   of the test strip and microneedle array holder. Once in
            studies with the 1 × PBS, 0.5 mg/mL histamine, and   place, a lateral flow test strip was placed  in  the
            1.0  mg/mL histamine solutions, nine test strips were   washed sample reservoir and 1000 µL of sample dilu-
                                                                                ®
            used  for  each type  of tuna  sample.  Five  test  strips   ent from the Reveal  kit was added to the inlet port
            were used  for the time course study. Four test strips   chamber of the device. The diluent traveled through
            for each method were used in the time course study.     the inlet channel, washed over the  microneedles that
               For the calibration studies and the spoiled fish study,   were used  to  sample the tuna flesh, and  entered  the
                                              ®
            the protocol [32]  described by the Reveal  for histamine   chamber holding the lateral flow test strip. Figure 2F
            test kits was used to validate the findings of the mi-  shows an example of a completed  test that followed
            croneedle sampling system procedure. Briefly, the   this procedure. The lateral flow strip is shown to ex-
            procedure  involved  acquiring  a  10  g  piece  of  tuna   tend vertically out from the reservoir chamber, having
            flesh, blending the tuna flesh in a food prep blender   been exposed to the diluent that flowed over the mi-
            until it was homogenized, and adding 190 mL of deio-  croneedles located in the central chamber. The lateral
            nized  water to the blended sample. The  sample  was   flow test strip in Figure 2F shows a positive histamine
            then hand-shaken for 20 seconds and allowed to rest   reading; a pink control band is observed in the upper
            for 5 minutes; the shaking and resting sequence was   portion and a more faintly pink test band is observed
            then  repeated. Immediately prior to  sampling  the   in the central portion. A ratio of the test: control line
            blended fluid, the container was shaken; the tuna sam-  color intensity  below 1  is indicative of a positive
            ple was allowed to settle in the suspension and 100 µL   reading.
            of fluid was removed. This 100 µL sample was added
            to a bottle containing sample diluent and then mixed.   3.2 Comparison of Testing Methods with Histamine-
                                                               Spiked Tuna Samples
            200 µL of fluid was removed from the sample diluent
            bottle provided in the test kit; this fluid was added to a   For  calibration  of the  microneedle sampling  system
            small sample cup, which contained a lateral flow test   procedure with the tuna flesh, pieces of tuna were in-
            strip. The lateral flow test strip was allowed to devel-  cubated overnight in solutions of 1 ×  PBS (negative
            op in the solution for at least 5 minutes and then eva-  control), 0.5 mg/mL histamine, and 1.0 mg/mL hista-
                                   ®
            luated  with  the Accuscan   reader. A positive result   mine. The  microneedle sampling system procedure
                             ®
            from the Accuscan  reader indicated that the sample   was used to obtain fluid from the tuna sample. The
            contained greater than 50 ppm of histamine.        same tuna sample was then evaluated using the
                                                               manufacturer-described procedure (i.e., the procedure
            3. Results and Discussion                          described in the Reveal  for Histamine kit). Compar-
                                                                                   ®
                                                               isons between the  data  acquired from both testing
            3.1 Microneedle and Test Holder Fabrication
                                                               methods for each  solution  type were  made using  a
            The microstereolithography process was successful in   difference of the means with a 95% confidence interval.

                                        International Journal of Bioprinting (2016)–Volume 2, Issue 1      75
   74   75   76   77   78   79   80   81   82   83   84