Page 176 - AJWEP-v22i3
P. 176

Mounira

                using  an  appropriate  factor.  BOD 5  values (mg/L) are   the  reading  was  allowed  to  stabilize  before  being
                calculated using Equation II.                       recorded.  The  instrument  was  calibrated  according

                BOD 5  = Reading × Correction factor          (II)  to  the  manufacturer’s  guidelines  before  use.  DO  was
                                                                    expressed in mg/L and as a percentage of saturation (%).
                2.5.5. COD measurement
                COD was measured by heating a diluted water sample   2.5.11. Nitrite measurement
                with potassium dichromate in a COD reactor at 150°C   Nitrite concentration was determined using a DR/2000
                for 2 h. A blank sample (reagent + distilled water) was   spectrophotometer  with  a  specific  program  (Program
                prepared  similarly.  After  cooling,  the  absorbance  of   345) (Hach DR2000, Germany). After sample preparation
                both the sample and the blank was measured at 420 nm   and reaction with a NitriVer 3 Nitrite Reagent Powder
                using  a  spectrophotometer  (DR2000,  Germany).  The   Pillows, the absorbance of the sample was measured at
                COD value was determined based on the difference in   507 nm with a blank used for zeroing calibration. Nitrite
                absorbance between the sample and the blank.        concentration was expressed in mg/L.

                2.5.6. Nitrate concentration measurement            2.6. Statistical analysis
                Nitrate concentration was determined using a DR/2000   All statistical analyses were performed using Microsoft
                spectrophotometer with a specific program 355 which   Excel. Descriptive statistics, including means, standard
                was set to 500 nm (DR2000, Germany). After dilution   deviations, maximum and minimum values, and 95%
                and reaction with a Nitraver 5 capsule, the absorbance   CI, were calculated  for all  measured parameters.  To
                of the sample was measured at 500 nm, with a blank   assess  the  statistical  significance  of  the  reduction  in
                used for calibration. The final result was adjusted for   COD concentrations between the influent and effluent
                the dilution factor.                                wastewater, a paired samples t-test was conducted.
                                                                    The paired  t-test  was  chosen  due  to  its  suitability
                2.5.7. Ammonium (NH 4 ) concentration measurement   for comparing means from two related  groups
                NH 4   concentration  was  measured  using  a  DR/2000   (i.e.,  measurements  taken  before  and  after  treatment
                spectrophotometer  with  a  specific  program  (DR2000,   from the same source).  A  p  < 0.05 was considered
                Germany) at 425 nm. After dilution and the addition of   statistically significant.
                stabilizing and Nessler reagents, the sample color was
                measured against a distilled water blank after a 1-min   3. Results and discussion
                reaction period.
                                                                    3.1. Temperature
                2.5.8. PO 4 concentration measurement               Temperature  was measured  using a thermometer  for
                         3−
                PO 4 concentration  was  measured  using  a  DR/2000   the area and both inflow and outflow wastewater. These
                   3−
                spectrophotometer  with  a  specific  program  (DR2000,   measurements are displayed in Figure 3 and Table 3.
                Germany) at 890 nm. After dilution and reaction with   From Figure 3, we can observe that the temperatures of
                a ChromaVer. 3 reagent capsule, the blue color of the   the inflow and outflow wastewater are relatively similar,
                sample  was  measured  against  a  distilled  water  blank   with the two curves nearly overlapping. The average
                following a 2-min reaction period.                  temperature of the inflow wastewater is 20.50 ± 3.97°C,
                                                                    while the average temperature of the outflow wastewater
                2.5.9. Temperature measurement                      is  20.76  ±  4.26°C.  The  maximum  temperatures  for
                Temperature  measurements  were  performed  in  situ   the  inflow  and  outflow  wastewater  are  26.75°C  and
                using a digital thermometer integrated into the portable   27.50°C, respectively.  The minimum temperature  for
                conductivity meter (WTW, Germany). The thermometer   the inflow wastewater is 11°C, while the minimum for
                probe was immersed directly into the water sample, and   the outflow wastewater is 12°C.
                the temperature reading was recorded once it stabilized.
                Temperature was expressed in degrees Celsius (°C).  3.2. pH
                                                                    The  pH  values  of  the  inflow  and  outflow  wastewater
                2.5.10. DO measurement                              were measured, and the results are presented in
                DO  concentrations  were  measured  in  situ  using  a   Figure  4 and  Table  4.  The results demonstrate  that
                portable DO meter (WTW, Germany). The DO probe      the  pH  values  of  both  the  influent  (mean:  7.82,
                was immersed  directly  into the  water  sample,  and   95%  CI:  7.74  –  7.90)  and  effluent  (mean:  7.95,  95%



                Volume 22 Issue 3 (2025)                       170                           doi: 10.36922/AJWEP025120085
   171   172   173   174   175   176   177   178   179   180   181