Page 189 - IJB-9-6
P. 189

International Journal of Bioprinting                    3D-printed assembly anatomical patella fracture bone plate




            Table 3. Results of the mean (standard deviation) gap distances between AATBP and each of the patella surfaces at the positions of
            the four vertices (P1, P2, P3, and P4)
                                  P1                   P2                    P3                   P4
             Sample 1             0.50                 0.41                  0.09                 1.01
             Sample 2             0.30                 0.73                  0.11                 1.14
             Sample 3             1.36                 0.16                  0.49                 0.57
             Mean ± Std (mm)      0.72 ± 0.46          0.43 ± 0.23           0.23 ± 0.18          0.91 ± 0.24
            Unit: mm

            surface was defined as a compression-only support in   Table 4. Results of the assembly tightness error between the
            analysis that enables the surfaces to only bear compression   PP and DP
            and not tension forces to mimic support from distal         Assembly tightness error between PP and DP
            femur condyles (Figure 9). The fractured gap distances at
            the medial and lateral sides for two fixation models and   Sample  Proximal side (mm)  Distal side (mm)
            maximum von Mises strain of bone around the screw hole   1         0.39            0.16
            were recorded.                                      2              0.31            0.15
                                                                3              0.33            0.20
            3. Results                                          Mean ± Std (mm)  0.34 ± 0.04   0.17 ± 0.03
            3.1. Design and 3D printing feasibility analysis   Unit: mm
            The mean (standard deviation) gap distance results
            between the AATBP and each of the patella surfaces at the   as 143.58 N (Figure 10). In contrast, the AATBP fixation
            positions of the four vertices (P1, P2, P3, and P4) of the   force–displacement curve was found to be smooth until
            15 mm × 15 mm square for the three patellae were 0.72 ±    tensile testing stopped. The dynamic test results showed
            0.46 mm, 0.43 ± 0.23 mm, 0.23 ± 0.18 mm, and 0.91 ±   that the average fractured gap distances at the medial/
            0.24 mm (Table 3). The AATBP manufacturing errors were   lateral sides before and after cyclic load testing were 2.38 ±
            both within 5% (maximum error of 3.75%), which showed   0.57 mm/2.30 ± 0.30 mm for TBW and 0.03 ± 0.01 mm/
            that the metal 3D printing equipment used in this study has   0.06 ± 0.03 mm for AATBP fixations (Table 5). This
            good precision and was suitable for medical applications   showed that the variations in the fractured gap distances
            (Table 1). The assembly tightness errors between the PP   were significantly different between TBW and AATBP,
            and DP at the proximal and distal positions were only 0.34   irrespective of whether the medial or lateral side was
            and 0.17 mm, respectively (Table 4).               analyzed, with all p < 0.05 using the Kruskal–Wallis test
                                                               because sample sizes were relatively small, and it could not
            3.2. AATBP bending strength and surface roughness  be determined whether the data were normally distributed.
            In the static four-point bending test, the values of the   Figure 11 shows all samples fixed with TBW/AATBP after
            average proof load and bending strength for AATBP were   dynamic testing to show the fracture gaps, and we found
            58.38 ± 4.51 N and 1167 ± 90.39 N·mm, respectively. The   no sustained damage on the AATBPs.
            failure mode of the AATBP after static tensile test showed
            that the plate was damaged around the screw holes because   3.4. Finite element analysis
            screws were pulled out from the rigid extension segment   The FE analysis result showed that the fractured gap distances
            (Figure 6b and c). The surface roughness test results showed   at the medial/lateral sides for TBW and AATBP fixation
            that the mean (standard deviation) surface roughness   models  were  0.8099  mm/1.1838  mm  and  0.3049  mm/
            for L1, L2, L3, and L4 segments were 4.42 ± 0.28 µm,    0.3136 mm, respectively. The AATBP fixation fracture
            5.59 ± 0.33 µm, 5.39 ± 0.91 µm, and 3.56 ± 0.25 µm,   gap was found obviously smaller than those of TBW
            respectively (Table 2).                            fixation under receiving tensile loads regardless of medial
                                                               or lateral sides. The maximum von Mises stress of the
            3.3. Biomechanical static/dynamic testing          bones were found around the medial wire hole and lateral
            The static test for TBW fixation found that the average   bottom locking screw hole for TBW and AATBP fixations,
            fractured gap distance  before and  after  the  cyclic  loads   respectively (Figure 9). The corresponding value for TBW
            was  3.06  ±  1.97  mm  (>2  mm),  and  the  corresponding   fixation was 17 MPa and much higher than that of 5.69 MPa
            force was 351.98 N. The average force of the region with   for AATBP fixation. Nevertheless, both bony stress values
            a volatility in the force–displacement curve was recorded   were a lot smaller than the bone fracture strength.

            Volume 9 Issue 6 (2023)                        181                         https://doi.org/10.36922/ijb.0117
   184   185   186   187   188   189   190   191   192   193   194